UVC/T-M-AR, DNA/RNA UV-Cleaner Box with built in socket, without table
Product code : BSN.BS-040104-A06
BSN.BS-040104-A06DNA/RNA UV-cleaner box UVC/T-M-AR is designed for clean operations with DNA samples. UV--cleaner box provide protection against contamination. Model is a bench-top type, made of metal framework, glass walls, working surface made of stainless st…
Via de knop “offerte aanvraag” kan u vrijblijvend contact met ons opnemen voor verdere informatie.
UVC/T-M-AR, DNA/RNA UV-Cleaner Box with built in socket, without table
Product beschrijving
DNA/RNA UV-cleaner box UVC/T-M-AR is designed for clean operations with DNA samples. UV--cleaner box provide protection against contamination.
Model is a bench-top type, made of metal framework, glass walls, working surface made of stainless steel.
UV-cleaner boxes are equipped with an open UV lamp installed in the upper hood. UV-radiation from the open lamps disinfects the working area inactivating DNA/RNA fragments during 15-30 min of exposure. A digital timer controls duration of the direct UV irradiation. A daylight lamp provides proper illumination of the working surface.
UV-cleaner boxes are equipped with a flow-type bactericidal UV cleaner-recirculator AR, which provides constant decontamination inside the box during operation. They are recommended for operations with DNA/RNA amplicons.
UV cleaner-recirculator AR consists of a UV lamp, a fan and dust filters organized in a special body so that a user working with a UV-cleaner box is protected against UV light. Recirculator increases the maximum density of UV light making it sufficiently effective for DNA/RNA inactivation. The UV-recirculator processes 100 UV-cleaner box volumes per hour, creating permanent aseptic conditions of operation inside the UV-cleaner box.
Specially assigned moving tables T-4 (with wheel locks) with a drawer are available on request.
Advantages of Biosan UV-cleaner boxes:
- Ozone free high density UV decontamination
- Long living UV lamps (9,000 hours average)
- Automatic switch off of UV-lamps when the protective screen is opened
- Bactericidal flow-type recirculator providing permanent decontamination inside UV -cleaner box during operation
- Shockproof glass walls
- Low noise, low energy consumption
- Tables for installation of UV-cleaner boxes
- UV-cleaner boxes with the bactericidal UV cleaner-recirculator AR is the patented Biosan solution
UVC/T-M-AR, DNA/RNA UV-Cleaner Box with built in socket, without table
Product specificaties
Packaging Unit
1 Unit
Brand
Biosan
Daylight lamp (for working area illumination)
1 x TLD-15W
Digital time setting of direct UV exposure
1 min-24 hrs / non-stop (increment 1 min)
Input current/power consumption (230V)
67 W
Nominal operating voltage
100-240 V, 50/60Hz
Open UV-lamp
1 x 25W built-in bactericidal, TUV25WG13 UV-C
Optical transmission
0,95
Optional table
T-4 (800 x 600 x 750 mm)
Overall dimensions (WxDxH)
690 x 535 x 555 mm
Radiation type
UV (λ = 253.7 nm), ozone-free
Safety
Automatic open UV-lamp switch off when screen is open
UV protection
>96%, UV-protection film
UV radiation level
15 mW / cm2 / sec
UV-recirculator
1x25W (efficiency >99% per 1 hour)
Walls material
Rear: stainless steel Sides and front: glass (EUROGLASS, Germany)
Weight
28.8 / 39 kg (net/gross)
Working area
650 x 475 mm
Working surface material
Stainless steel